2. Technical System Description 1/2010 2 - 22
BA-TE-DE08C M.KAY
Dialog+ SW9xx_sm_Chapter 2_1-2010.doc/pdf <100329> yymmdd
B. Braun Avitum AG
Dialog
+
SW 9.xx
2.18 Level Regulation Module
V
P
E
V
P
D
V
P
U
V
P
V
V
P
A
V
B
T
P1
P2
P3
LRP
HFS
HFE
HFV
HFA
PA
PBS
PV
PBE
HFP
PA PV PBS PBE
P
OUT
P
IN
HFS
HFE
HFV
HFA
PA
PBS
PV PBE
LRP
P
OUT
P
IN
Fig. : Level Regulation Module and Level Regulation Pump
2.18.1 Legend Level Regulation Module
There are two level regulation modules with different assemblies, i.e. for machines with a single blood pump and with a
double blood pump.
Inlet Pressure Sensor Valve VPE
VPEVPE
VPE
Down Pressure Sensor Valve VPD
VPDVPD
VPD
Up Pressure Sensor Valve VPU
VPUVPU
VPU
Blood Control Pressure Sensor Valve VPS
VPSVPS
VPS (Double Pump)
Venous Pressure Sensor Valve VPV
VPVVPV
VPV
Arterial Pressure Sensor Valve VPA
VPAVPA
VPA
Blood Side Test Pressure Sensor Valve VBT
VBTVBT
VBT
Level Regulation Pump LRP
LRPLRP
LRP
Arterial Hydrophobic Filter HFA
HFAHFA
HFA
Venous Hydrophobic Filter HFV
HFVHFV
HFV
Blood Control Pressure Hydrophobic Filter HFS
HFSHFS
HFS
(Double Pump)
Inlet Hydrophobic Filter HFE
HFEHFE
HFE
Pump Hydrophobic Filter (Ceramic Filter) HFP
HFPHFP
HFP