CAPINTEC, INC. CAPTUS
®
3000
Figure 11-7
vi. The screen shown in Figure 11-8 appears next, prompting you to measure the
selected isotope for efficiency measurement.
Figure 11-8
To measure the isotope, place the isotope standard in the Neck Phantom exactly
simulating the geometry used in the bioassay measurement and click the
button. Data collection begins as shown in Figure 11-7 above. To
stop the measurement, click the
button. The screen then appears as
shown in Figure 11-9.
Figure 11-9
11-6 BIOASSAY November 13