NIBP Monitoring
12-7
2) Select [Setup] tab.
3) Set [Start Mode] to [Clock].
4) Select the drop-down list on the right of [Interval] and set the specific measurement interval.
5) Select [NIBP Start/Stop] hot key or select [Start NIBP] in [NIBP] menu to start the first measurement.
Or you can select [Clock], and then [NIBP Setup] → [Interval] to start the first measurement.
After the first measurement is completed the monitor will automatically repeat the measurement with the set
interval. For example, if the first measurement begins at 08:23, and the interval is set to [5 min], the monitor will
automatically start the next measurement at 08:25. Subsequent NIBP measurements are made synchronously with
the clock, then at 08:30, and so on.
12.7.2.4 Start NIBP STAT
You can start NIBP STAT in any of the following ways:
Select [NIBP STAT] hot key.
Select [NIBP Setup] hot key → [NIBP STAT].
Enter [NIBP] measure → select [NIBP STAT].
The NIBP STAT process lasts for 5 mins.
12.7.2.5 Start Sequence Measure
1) Select the NIBP parameter area to enter [NIBP] menu.
2) Select [Setup] tab.
3) Set [Interval] to [Sequence].
4) Set [Start Mode] to [Interval].
5) Set parameters for the sequence; please refer to "
Section 12.6.5 Set NIBP Measure Sequence
" for more
information.
6) Select [NIBP Start/Stop] hot key or enter [NIBP] menu → select [Start NIBP] or [NIBP Setup] hot key → select
[Sequence] to start the first measurement.
After the first measurement is completed the monitor will start measurement automatically in the times and
intervals of the set sequence.