Alarms
7-8
7.9.5 Alarms All Off
If [Pause Time] is set to [Permanent] (refer to “
Section 7.9.3 Set Alarm Pause Time
”), the [Alarm Pause] hot key is
changed to [Alarm off] key. All alarm signals will be shut down if [Alarm Pause] hot key is pressed, that is, alarms
are off, and:
the light and alarm audio of alarm conditions will be cancelled.
[Alarm Off] with red background will appear in the physiological message area.
the alarm off icon will be shown in the message area.
Click this hot key again to exit the alarm off status.
Risk may exist if the alarm function is paused or turned off. The user should pay close attention to the
actual clinical conditions of patients.
7.9.6 Alarm Audio Pause
The user can pause the alarm sound by pressing the [Alarm Audio Pause] hot key. When the alarm sound is in the
paused state:
The alarm audio of physiological alarm and technical alarm conditions will be muted in the specified time.
The remaining time of alarm audio paused will appear in the physiological alarm message area.
The alarm audio paused icon will be displayed in the message area.
The monitor will exit the alarm audio paused status automatically when the specified pause time expires. You can
also click [Alarm Audio Pause] hot key to cancel the alarm audio paused status.
7.9.7 Set Alarm Audio Pause Time
Set alarm audio pause time: [1min], [2min], [3min], [Permanent]. Set alarm audio paused time as shown in "
Section
7.9.3 Set Alarm Pause Time
".
7.9.8 Delay Alarm Audio Pause Time
When the monitor is in the alarm audio paused status, you can increase the audio paused time temporarily.
Increase alarm sound pause time as shown in "
Section 7.9.4 Delay alarm pause time
".