Alarms
7-14
7.19 Intubation Mode
The monitor provides an intubation mode when monitoring the parameters of RESP, CO
2
, AG and RM. Intubation
mode is a unique monitoring mode. In this mode, the parameters of RESP, CO
2
, AG and RM are all blocked, and the
alarm off icon appears in the parameter area. This mode is suitable to be used under general anesthesia where
there is a need to inhibit unnecessary alarm conditions.
7.19.1 Enter Intubation Mode
The user can set the monitor to intubation mode by the following methods:
Click [Intubation Mode] hot key;
Click [Intubation Mode] button under the menu of [RESP], [CO
2
], [AG] or [RM]; or
Click [Main Menu] hot key → [Alarm] → [Intubation Mode].
7.19.2 Set Intubation Mode Time
1) Click [Main Menu] hot key → [System] → [Maintenance], enter the maintenance password and click the Enter
key.
2) Select [Alarm] tab → [Other] tab.
3) Set [Intubation Mode Time].
7.19.3 Exit Intubation Mode
The user can exit intubation mode by the following methods:
Click [Exit Intubation Mode] hot key;
Click [Exit Intubation Mode] button under the menu of [RESP], [CO
2
], [AG] or [RM]; or
Click [Main Menu] hot key → [Alarm] → [Exit Intubation Mode].
7.20 Alarm Auto Print
You can set the waveform printing duration when an alarm occurs as follows:
1) Enter [Setup] screen by any of the following methods:
Click [Alarm Setup] hot key.
Click [Main Menu] hot key → [Alarm] → [Config.].
2) Select [Setup] tab.
3) Set [Printing Duration On Alarm].