Alarms
7-15
7.21 Restore Alarm Defaults
1) Click [Main Menu] hot key [System] [Maintenance], enter the maintenance password and click the Enter
key.
2) Select [Alarm] tab.
3) Press the [Defaults] button.
7.22 Alarm System Self-test
Upon startup of the device the alarm system will perform a self-test of the alarm light and alarm audio.
Alarm lights and alarm sound characteristics during self-test are:
The red alarm light is on for 1s; then the yellow light and the cyan light are on for 1s respectively;
The alarm system makes a “beepsound as the alarm sound self-test.
If further testing of alarm system is needed, you can use an appropriate simulator to test the measurement units,
and adjust the alarm limits to verify whether the correct alarm response can be triggered.