EasyManua.ls Logo

Comen N10 - Standby Mode; Demo

Default Icon
307 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Basic Operations
4-4
You can exit Privacy Mode:
by selecting [Exit Privacy Mode] on the screen; or
by disconnecting from the CMS; or
when the battery level is too low.
4.2.4 Standby Mode
If patient monitoring is not needed for the time being, but you don't want to shut down the device, Standby Mode
can be used.
Here are the steps for entering Standby Mode:
1) Select the [Standby] or [Main Menu] hot key select [Standby] from the [Patient Management] column.
2) Set the patient location in standby.
3) Select OK.
Performance of the device in standby mode:
Stops all parameter measurements.
All alarm messages (except Low Battery) and prompt messages are shielded.
Exit Standby Mode:
Select [Resume Monitoring] to exit standby mode and restore monitoring for the current patient.
Select [Discharge Patient] to discharge the current patient from the monitor.
NOTE
When exiting Standby Modethe alarm of [Alarm Paused] will be enabled for two minutes.
4.2.5 Demo
Here are the steps for entering Demo Mode:
Select [Main Menu]; select [Demo] from the [System] column; input the password to enter Demo Mode.
CAUTION
Demo waveforms are simulated waveforms set by the manufacturer only for the purpose of
demonstrating device performance and aids in user training. In actual clinical use, it is forbidden to use
the demo function because medical workers may mistake the demo data for waveforms and parameters
of the patient being monitored, which will affect patient monitoring and delay diagnosis and treatment.

Table of Contents