Alarms
7-4
Alarm reset: indicates that alarms are acknowledged and the alarm system is reset.
7.5 View Physiological Alarm conditions
View physiological alarms as follows:
1) Select physiological alarm area to enter [Alarm] window.
2) Select [Physiological Alarms] Tab. The current alarms are in the displayed list.
3) Select [Details] to get more information.
7.6 View Technical Alarm conditions
View technical alarms as follows:
1) Select technical alarm area to enter [Alarm] window.
2) Select [Technical Alarms] tab. The current alarms are in the displayed list.
3) Select [Details] to get more information.
7.7 Set Alarm Limits
7.7.1 Manual Adjust Alarm Limit
1) Enter [Limits] screen in any of the ways listed below:
Click [Alarm Setup] hot key.
Click [Main Menu] hot key [Alarm] [Limits].
2) Set the alarm parameters in corresponding tabs as required.
You can also set the alarm limit for one parameter from the parameter menu.
7.7.2 Set Auto Alarm Limits
The monitor provides an Auto Alarm Limit Setup function. This function allows automatic setup of alarm limits
through the currently measured parameter values based on the current patient type. These alarm limits will remain
until you initiate [Auto Limit] or manually adjust alarm limits.
Before applying these alarm limits, please confirm whether they are appropriate for the current patient. If not, you
will need to manually set the alarm limits.
1) Enter [Limits] screen in any of the ways listed below:
Click [Alarm Setup] hot key.
Click [Main Menu] hot key → [Alarm] → [Limits].