CO
2
Monitoring
15-20
WARNING
Replace the airway adapter if rainout/condensation occurs inside the airway adapter.
The IRMA airway adapter is designed as a non-sterile single patient use disposable for both Adult/Pediatric and
Infant applications. The IRMA Infant airway adapter has specially designed connectors for minimizing the dead
space and can be used even for very small patients.
IRMA airway adapters: Adult/Pediatric (REF: 106220) and Infant (REF: 106260)
WARNING
Do not use the IRMA Adult/Pediatric airway adapter with infants as the adapter adds 6 ml dead space
to the patient circuit.
Do not use the IRMA Infant airway adapter with adults as this may cause excessive flow resistance.
15.9.7 Maintenance
The user should verify gas readings regularly; If any problems occur, contact an engineer of the manufacturer for
maintenance