21.6 Set the Printed Report
21.6.1 Set the ECG Report
1) Click [Main Menu] hot key → [Report] →[Normal Report].
2) Click [ECG Report].
3) Set the desired menu items. Part of menu items are described in the table below.
Set the amplitude of ECG
waveform
To record and print out the ECG waveform in that
amplitude
Set the print speed of ECG
waveforms
[25 mm/s]: prints 25 mm of ECG waveform per second.
Set the format of the 12-lead
output waveform
[12×1]: print the 12 waveforms on one piece of paper
from top to bottom.
You will fail to print the ECG report if [Lead Type] is set to [3 Leads].
21.6.2 Set the Realtime Report
1) Click [Main Menu] hot key → [Report] → [Normal Report].
2) Click [Realtime Report].
3) Set the desired menu items. Part of menu items are described in the table below.
Set the print speed of ECG
waveforms
[Auto]: Automatically print the waveform
[12.5 mm/s]: print 12.5 mm of ECG waveform per second
[25 mm/s]:print 25 mm of ECG waveform per second
[50 mm/s]:print 50 mm of ECG waveform per second
Choose the waveforms to be
output
[Current Waveforms]: output the waveforms being
displayed on the screen.
[Selected Waveforms]: print the realtime report of
selected waveforms.
21.6.3 Set the Tabular Trends Report
1) Click [Main Menu] → [Report] → [Normal Report].