Maintenance
24-2
24.2 Maintenance Plan
The following tasks can be fulfilled only by professional maintenance personnel recognized by the Company. If the
following maintenance is required, please timely contact the maintenance personnel. Prior to test or maintenance,
the device must be cleaned and disinfected.
Test and Maintenance Items
Frequency
Visual inspection and performance inspection
Visual inspection
Before first use every day
Parameter module performance inspection
and calibration
1. When measurement inaccuracy is suspected
2. When the relative module is maintained or replaced.
3. At least once every year for CO
2
module and AG module
4. At least once every two years for other modules
Analog output test
As required
Defibrillation Synchronization test
As required
Nurse Call test
As required
Safety inspection
Perform safety inspection according to
GB9706.1/IEC 60601-1
1. When the power module is maintained or replaced
2. If the monitor has been dropped
3. At least twice every year or as required
Other inspections
When the device starts
Each time the monitor is used.
Recorder test
Before first use
When the recorder is maintained or replaced
Network printing test
After the initial installation
When the printer is maintained or replaced
System integrated test
1. After the initial installation
2. When the integrated device is maintained or replaced
Battery inspection
Function test
1. After the initial installation
2. After the battery is replaced
Performance test
Every three months or the battery running time has shortened
noticeably.
NIBP leakage test
At least once every two years or as required.
NIBP verification
At least once every two years or as required.
ECG calibration
At least once every two years or as required.
IBP calibration
At least once every two years or as required.
Mainstream and sidestream CO
2
calibration
and performance check
At least once every two years or when measurement inaccuracy
is suspected.
Battery
Refer to the battery-related chapter in this manual.