The Net Bed on the monitor must be unique to avoid conflicts with other Net Beds connecting to the
central monitoring system.
Refer to the instruction manual of our central monitoring system for details.
[Time Setup] on the monitor will turn grey and cannot be operated when it is connected to the central
monitoring system.
25.3 Connection to the Hospital's Information System by HL7 protocol
1) Click [Main Menu] hot key → [System] → [Maintenance], enter the maintenance password and click the Enter
key.
2) Select [Network Setup] tab → [HL7 Setup] tab.
3) Turn on the switch of [Send Data], [Send Waveforms], [Send Alarms] as required.
4) Select [IP Address] and [Port] to set the IP address and port of the server that receives real-time data and
waveforms.
5) Set [Data Interval].
You can view the server connection status through this page.
25.4 Remote View
Under the same network, Remote View function allows the monitor can observe all the measured parameters and
the alarms on other monitors or devices from a remote site or bed. The N monitor can simultaneously watch up to
18 remote devices.
25.4.1 Enter Remote View window
Enter [Remote View] screen in any of the ways listed below:
Press [Remote view] hot key.
Select [Screen Setup] hot key → [Primary Screen] tab → select [Choose Screen] tab → [Remote View.]