Installation and Preparation
3-2
Note
Condensation is the condensation of gas or liquid when it is cold, such as water vapor changing into
water when it is cold, and water changing into ice when it is cold. The lower the temperature, the faster
the condensation speed.
3.2 Device Preparation
3.2.1 Connection of the AC power supply cord
Before connecting the AC power supply cord, ensure that the AC supply voltage and frequency are consistent with
the voltage and frequency indicated on the device.
Steps for connecting the AC power cord:
1) Use the power cord supplied with the monitor; connect one end of the power cord to the inlet of the monitor;
2) Insert the other end of the power cord to a mains socket outlet with protective earth.
3) Confirm that the AC power indicator turns on, which indicates that AC power supply is connected normally.
NOTE
Connect the power cord to hospital-level outlet.
When a battery is provided, the battery must be charged after transportation or storage. If the battery
is low, the monitor may fail to work without connecting an AC power supply. Once the monitor is
connected to an AC power outlet, the battery will be charged whether the monitor is switched on or not.
Connect an equipotential grounding wire if necessary.
3.2.2 Connect to DC Power Source (for N10/N10A/N10C)
When it is used in an ambulance, the monitor N10/N10A/N10C can be connected to the vehicle DC power source:
1) The monitor is equipped with a DC power cord. Connect one end of the power cord to the DC power outlet on
the vehicle.
2) Connect another end of the power cord to the DC power inlet on the monitor
3) Confirm that the power indicator lights on, which suggests that the DC power source is connected correctly.
WARNING
It is only allowed to use the DC power cord equipped with the monitor. Do not use AC/DC power
adapter.
When using DC power supply, do not touch the patient and the mental parts or sockets of the monitor
at the same time.
Before connecting to DC power supply, confirm whether the DC voltage and frequency satisfy the
specifications specified on the manual or device label.