Alarms
7-10
2) Select [Alarm] tab→[Other] tab
3) Switch ON or OFF the [CMS Disconnection Alarm]: if OFF is selected, the alarm information [CMS Disconnected]
will not be generated when the monitor is disconnected from the CMS.
7.13 Set the Switch Status of CMS Alarm System Control
User can choose whether to allow the CMS monitoring system to control the Monitor’s alarm system by:
1) Select [Menu] hotkey→[System]--- [Maintenance]→input password→click Enter key
2) Select [Alarm] tab→[Other] tab
3) Switch ON or OFF the [CMS Alarm System Control].
7.14 Set Alarm Delay Time
Alarm delay time can be set for those continuously monitored parameters. The monitor would not activate alarm
signals if within the alarm delay time the specific conditions that trigger the alarm no longer exist. Set as follows:
1) Click [Main Menu] → [System] → [Maintenance], enter the maintenance password and click the Enter key.
2) Select [Alarm] tab → [Other] tab.
3) Set [Alarm Delay].
This alarm delay time does not apply to HR-related alarm conditions.
7.15 Alarm Reset.
The monitor provides Alarm reset control to confirm the current alarm conditions, and reset the alarm system (the
alarm system is enabled).
Click [Alarm Reset] hot key, the alarm reset icon is displayed in the message area.
If a new alarm is generated when the monitor is in alarm reset state, the alarm reset icon will
disappear and the sound/light of the alarm will function normally
7.15.1 Physiological Alarm Reset
When the physiological alarm is reset,
The sound of the existing physiological alarm condition will be silenced.
A symbol will appear before the alarm message, indicating that the alarm has been acknowledged.
The color of the parameter numeric background keeps flashing.