C.O. Monitoring
16-4
3) Set [Interval (s)]: the minimum time interval between two measurements; the next measurement will not be
carried out until the set interval is reached.
16.5.6 Set C.O. Coefficient
1) Select the C.O. parameter area to enter [C.O.] menu.
2) Select [Setup] tab.
3) Set [C.O. Coefficient]: The C.O. Coefficient is related to the Swan-Ganz catheter and injected volume. It should
be adjusted according to the manufacturer’s instructions when the catheter is replaced.
16.5.7 Set Injection Temp
1) Select the C.O. parameter area to enter [C.O.] menu.
2) Select [Setup] tab.
3) Set [Injection Temp]:
If [Auto IT] is set to On, the injection temperature will be measured automatically.
If [Auto IT] is set to Off, you need to manually input the injection temperature at [IT(℃)].
16.5.8 Set Inject Volume
1) Select the C.O. parameter area to enter [C.O.] menu.
2) Select [Setup] tab.
3) Set [Inject Volume (ml)].
16.5.9 Set Measure Mode
Measure Mode defaults to [Manual Measure], but you can set it to Auto Measure in the following way:
1) Select the C.O. parameter area to enter [C.O.] menu.
2) Enable [Auto Measure]. The C.O. measurement will start automatically at the set [Interval], and there is no
requirement to select the [Start] button. If [Start] is switched to [Stop], please start the injection and
measurement immediately.
3) If [Auto Measure] is disabled, you need to select [Start] for each measurement.
16.6 Cardiac Output Measurement