Basic Operations
4-2
4.1.4 Use of a Scanner
This monitor supports barcode and QR code scanners which can be connected to the monitor via a USB port.
NOTE
The QR code scanner customized by Comen can be used to scan QR codes and barcodes. Other QR code
and barcode scanners used can only read medical record numbers or registration numbers.
4.1.4.1 Clear Data Format
If you use the QR code scanner customized by Comen, it is necessary to clear the previous data format and configure
the scanner before initial use. Here are the steps for clearing data format:
1) Scan the barcode or engineering QR code for format clearing to clear the old data format.
2) Scan the engineering QR code used by the hospital to obtain the QR code format specifically for this hospital.
NOTE
Please contact the scanner manufacturer or Comen to obtain the barcode or engineering QR code for
format clearing.
4.2 Working Modes
4.2.1 Monitor Mode
Monitor Mode is used for patient monitoring. Upon startup of the device, it will automatically enter monitor mode.
4.2.2 Night Mode
This monitor provides a Night Mode with night-mode settings that can avoid disturbing the patient in the nighttime.
When exiting night mode, the monitor will also exit night-mode settings and restore the settings of other modes.
In Night Mode, the monitor uses the following settings by default:
Brightness: 1
Alarm Volume: 2
QRS Volume: 1
Key Volume: 0
NIBP End Tone: OFF
Stop NIBP Measure: OFF
Activate night mode: