Battery
22-3
must be performed every three months. When you suspect a battery fault, you will also need to perform a battery
performance check.
For the battery performance check procedure, see steps 1 to 4 in
Section 22.4.1 Optimize Battery Performance”
.
The length of discharge time reflects the performance of the battery. If the battery's power supply time is
significantly lower than the time stated in the Specifications, the battery should be replaced.
NOTE
In order to prolong the service life of the rechargeable battery, if the battery is stored for a long period
of time it is suggested that the battery should be charged every three months to prevent excessive
discharging.
The maximum recommended number of battery charge and discharge cycles is 300.
The voltage supply time of the battery depends on the configuration and operation of the device. For
example, frequent NIBP measurement will reduce the voltage supply time of the battery.
22.5 Battery Recycling
Discard the battery if it has signs of damage visually or the battery is aged or fails. Waste batteries should be properly
recycled in accordance with applicable laws and regulations or the rules of the hospital.
WARNING
Do not disassemble or short-circuit the battery or place it in fire; otherwise battery fire, explosion,
leakage of hazardous gas or other hazards may be caused.
Prior to disposal of the monitor, remove the battery. Then dispose of the device and its accessories in
accordance with your country’s regulations for equipment containing electronic parts.