Default configuration information
IV-2
2. Arrhythmia default settings
Arrhythmia Alarm default settings
Name
Alarm
Alarm Level
Alarm Outputs
R on T
Off
Med
Off
Bigeminy
Off
Med
Off
Trigeminy
Off
Med
Off
Nonsus V-Tach
Off
Med
Off
Vent Rhythm
Off
Med
Off
PVCs/min High
Off
Med
Off
Asystole
On
High
Off
V-Fib/V-Tach
On
High
Off
V-Tach
On
High
Off
Vent Brady
On
High
Off
Extreme Tachy
On
High
Off
Extreme Brady
On
High
Off
Run PVCs
Off
Low
Off
Couplet
Off
Prompt
Off
PVC
Off
Prompt
Off
Pacer Not Capture
Off
Prompt
Off
Pacer Not Pacing
Off
Prompt
Off
Missed Beats
Off
Prompt
Off
Irr Rhythm
Off
Prompt
Off
Tachy
Off
Med
Off
Brady
Off
Med
Off
Heart Pause
Off
Low
Off
Multiform PVC
Off
Med
Off
A-Fib
Off
Prompt
Off
Pauses/min High
Off
Med
Off
Irr Rhythm End
Off
Prompt
Off
A-Fib End
Off
Prompt
Off
Arrhythmia threshold default settings
Name
Patient type
Default settings
Asystole Delay (s)
ADU
5
PED
NEO
Heart Pause (s)
ADU
2.0
PED
NEO
Extreme Tachy (bpm)
ADU
160
PED
180
NEO
220
Extreme Brady (bpm)
ADU
35
PED
50
NEO
60