Default configuration information
IV-3
Tachy (The HR upper limit)
(bpm)
ADU
120
PED
160
NEO
200
Brady (The HR lower limit)
(bpm)
ADU
50
PED
75
NEO
100
Ventricular bradycardia
Rate (bpm)
ADU
40
PED
NEO
Ventricular bradycardia
PVCs
ADU
5
PED
NEO
V-Tac Rate (bpm)
ADU
130
PED
130
NEO
160
V-Tac PVCs
ADU
6
PED
NEO
Multi PVCs Window
ADU
15
PED
NEO
PVCs/min
ADU
10
PED
NEO
A-Fib/Irr Rhy End Time (s)
ADU
2
PED
NEO
Pauses/min
ADU
8
PED
NEO
3. ST default settings
Name
Default settings
ST Alarm Mode
Real-time value
ST-I, ST-II, ST-III, ST-aVR, ST-aVL, ST-aVF, ST-V1, ST-V2, ST-
V3, ST-V4, ST-V5, ST-V6, ST-Va, ST-Vb ([Alarm mode] is set
to [Real-time value])
Alarm On/Off
Off
Alarm Limits
(-0.2mV, 0.2mV)
Alarm Level
Med
Alarm Auto Print
Off
Single ST Alarm; Multi ST Alarm ([Alarm mode] is set to
[Various])
Alarm On/Off
Off
Alarm Limits
(-0.1mV, 0.1mV)
Alarm Level
Med
Alarm Auto Print
Off
ST Segment Analysis
Off
ST Segment
Auto