Default configuration information
IV-4
Show Markers
Off
ST Point
J+60ms
ST Point Adjustment
On
ISO Point
-80ms
J Point
48ms
4. QT/QTc default settings
Name
Default settings
QTc
Alarm On/Off
Off
Alarm Upper Limit (ms)
ADU: 500
PED: 480
NEO: 460
Alarm Level
Med
Alarm Auto Print
Off
QTc
Alarm On/Off
Off
Alarm Upper Limit (ms)
60
Alarm Level
Med
Alarm Auto Print
Off
QT Analysis
Off
QT Leads
All
5. Resp default settings
Name
Default settings
RR
Alarm On/Off
On
Lower/Upper Alarm Limits (bpm)
ADU: 8-30
PED: 8-30
NEO: 30-100
Alarm Level
Med
Alarm Auto Print
Off
Apnea
Alarm On/Off
On
Alarm Level
High
AlarmAuto Print
Off
Apnea Timeout
ADU: 20s
PED: 20s
NEO: 15s
RR Source
Auto
RESP Lead
ADU: Auto
PED: Auto
NEO: II
Gain
X1
Waveform Speed
12.5mm/s
Filter
On
Auto Threshold Detection
On