Clinical Assistant Assessment (CAA)
18-14
18.6.4 Start CCHD Screening
If the monitor you purchased is configured with two kinds of SpO
2
module (KPM module and SpO
2
plug-in module),
you can measure the SpO
2
in the right hand and foot simultaneously to detect CCHD. If only one SpO
2
module is
configured, you can measure the SpO
2
on the right hand first and then change to the foot. This can be done in the
following way:
1) Enter CCHD screening, see
Section 18.6.3 “Access to CCHD Screening
for detailed information.
2) Measure SpO
2
values
3) Check if the SpO
2
values displayed correspond to the measurement sites. If not, click the icon to record
the SpO
2
in right hand or foot.
4) Select heart murmur level if the [Two Standards] screening rule is applied,
5) Select [Accept] to acknowledge current measurement results and complete a CCHD screening.
The SpO
2
measurement result and the CCHD screening result are displayed on the CCHD screening window:
“Positive”: the CCHD screening is completed.
“Negative: the CCHD screening has failed and Echocardiography is strongly recommended.
“Repeat screen in 1 hour(American Standard: the CCHD screening is not confirmed and the user should
repeat the measurement in one hour.
“Repeat screen in 2-4 hour(Two Standards): the CCHD screening is not confirmed and the user should
repeat the measurement after 2-4 hours.
NOTE
During the measurement of SpO
2
, keep the newborn calm.
18.6.5 Screening again
If you have doubts about the screening result, please repeat the CCHD screening again. This action will clear the
measurement result and the screening result.
Select [Repeat CCHD] button, click [Yes] to start again.