Clinical Assistant Assessment (CAA)
18-14
18.6.4 Start CCHD Screening
If the monitor you purchased is configured with two kinds of SpO
2
module (KPM module and SpO
2
plug-in module),
you can measure the SpO
2
in the right hand and foot simultaneously to detect CCHD. If only one SpO
2
module is
configured, you can measure the SpO
2
on the right hand first and then change to the foot. This can be done in the
following way:
1) Enter CCHD screening, see “
Section 18.6.3 “Access to CCHD Screening
” for detailed information.
2) Measure SpO
2
values
3) Check if the SpO
2
values displayed correspond to the measurement sites. If not, click the icon to record
the SpO
2
in right hand or foot.
4) Select heart murmur level if the [Two Standards] screening rule is applied,
5) Select [Accept] to acknowledge current measurement results and complete a CCHD screening.
The SpO
2
measurement result and the CCHD screening result are displayed on the CCHD screening window:
“Positive”: the CCHD screening is completed.
“Negative” : the CCHD screening has failed and Echocardiography is strongly recommended.
“Repeat screen in 1 hour” (American Standard): the CCHD screening is not confirmed and the user should
repeat the measurement in one hour.
“Repeat screen in 2-4 hour” (Two Standards): the CCHD screening is not confirmed and the user should
repeat the measurement after 2-4 hours.
18.6.5 Screening again
If you have doubts about the screening result, please repeat the CCHD screening again. This action will clear the
measurement result and the screening result.
Select [Repeat CCHD] button, click [Yes] to start again.