Basic Operations
4-3
1) Select the [Main Menu] hot key → select [Night Mode] from the [Display] column.
2) Change the default settings of Night Mode.
3) Select [Enter Night Mode] to enter this mode.
Before entering Night Mode, please confirm the settings of Brightness, Alarm Volume and QRS Volume.
Please pay attention to potential risks if the settings are too low.
Exit Night Mode:
1) Select the [Exit Night Mode] or [Main Menu] hot key → select [Exit Night Mode] from the [Display] column.
2) Select [OK].
In Night Mode, if the monitor has been connected to the central monitoring system, it will automatically
exit Night Mode when disconnected from the central monitoring system.
4.2.3 Privacy Mode
When protection of the patient monitoring screen is required, Privacy Mode can be activated. This mode can be
activated only under CMS (Central Monitoring System) monitoring. Monitoring data is only displayed on the central
monitoring system.
You can enter Privacy Mode in either of the following ways:
Select the [Privacy Mode] hot key; or
Select the [Main Menu] hot key → select [Privacy Mode] from the [Display] column.
Performance of the device when Privacy Mode is activated:
The device interface is turned off, and the display of all parameters, waveforms and alarm messages is shielded.
Patient monitoring continues; patient data is saved but can be seen only in the central monitoring system.
Heartbeat sound and pulse sound from the monitor are shielded.
When Department is set to OR, Privacy Mode is not supported.
When the monitor generates the Low Battery alarm, it is impossible to enter Privacy Mode.
Privacy Mode is unavailable for N15C/N12C/N10C monitor.