To ensure normal operation of the monitor, please carefully read the contents of this chapter and the
“Safety Information” chapter prior to use, and install the monitor according to requirements.
This device shall be installed by personnel designated by our company.
3.1 Installation
3.1.1 Unpacking and Checking
Carefully take the monitor and its accessories out of the packing box and check according to the following aspects.
1) Check whether all accessories are provided according to the Packing List.
2) Check for damage.
3) Check all exposed lead wires and connectors.
4) For any problem or inconsistencies, contact Comen or your distributor.
5) Keep the packaging materials for future use.
3.1.2 Environmental Requirements
The operating environment for this device must conform to the environmental requirements specified in this
manual; otherwise the accuracy of the device may be affected, and damage to components and circuits may be
caused.
The patient monitoring system should be used in an environment that can reasonably avoid vibration, dust,
corrosive or explosive gases, extreme temperature and humidity, etc. Operating environment of this equipment
must meet the environmental specifications in this manual.
When the monitor is installed in an enclosed space, make sure the space is well ventilated. Leave at least 2 inches
(5cm) free space around the monitor for air circulation. Also, leave sufficient space around it for easy operation and
maintenance.
The device is not intended for hyperbaric operation. Use in a hyperbaric chamber can result in harm to the patient
and/or damage to the device.
Ensure that the monitor is free from condensation during operation. When the monitor is moved from one room
to another, condensation may be formed due to exposure to damp air and temperature difference. In this case, do
not use the monitor until it is dry.