NIBP Monitoring
12-8
12.7.3 Stop NIBP measurement
12.7.3.1 Stop Ongoing Measurement
You can stop the ongoing NIBP measurement in any of the following ways:
Select [NIBP Start/Stop] hot key.
Select the NIBP parameter area to enter [NIBP] menu select [Stop NIBP].
12.7.3.2 Stop NIBP STAT
You can stop NIBP STAT in any of the following ways:
Select [NIBP Start/Stop] hot key.
Select [NIBP STAT] hot key.
Select the NIBP parameter area to enter [NIBP] menu select [Stop NIBP].
Select the NIBP parameter area to enter [NIBP] menu select [Stop All].
12.7.3.3 Stop All NIBP Measurement
You can stop all NIBP measurements in any of the following ways:
Select [Stop All] hot key.
Select the NIBP parameter area to enter [NIBP] menu select [Stop All].
12.8 Venipuncture
Inflate the NIBP cuff to a pressure approximate to the diastolic pressure to block the vein vessel and assist in the
completion of venipuncture. Steps are as follows:
1) Select the NIBP parameter area Select [Setup] tab.
2) Set [Venipuncture Pressure].
3) Select [Venipuncture] button on the lower right corner of the menu or press [Venipuncture] hot key at the
bottom of the monitor.
4) Puncture the vein and take the blood sample.
5) Select [NIBP Start/Stop] hot key to deflate the cuff manually. If manual deflation is not performed, the
venipuncture process lasts for 125s for adult and pediatric patients and 87s for neonate patients. The cuff
deflates automatically after the said time period.
In the venipuncture process the NIBP parameter area will display the cuff pressure and remaining time of
venipuncture.