6-15
Chapter 6 Manage Configurations
6.1 Summary
For continuous monitoring of a patient, the user/operator may be required to adjust some settings according to the
actual patient conditions. The collection of pre-set items used to operate the monitor is called a configuration. In
order to configure the monitor more effectively and rapidly, this monitor provides a variety of configurations to
meet the requirements of different patient types and different hospital departments. You can also customize a
configuration according to actual conditions and save it as a user-defined configuration.
Configuration info of the monitor mainly includes:
Parameter Configuration
Setting items related to parameter measurements, such as Wave Gain, Speed, Unit, Alarm ON/OFF, and Alarm Limit
Setup.
General Configuration
The monitor’s general setting items, such as Alarm Setup, Screen Layout, and Record.
Maintenance Configuration
Settings related to user maintenance.
Department options: General (General Monitoring)
OR (Operation Room/Anesthesia Monitoring)
ICU (Intensive Care Unit)
Neo (Neonatal Intensive Care Unit)
CCU (Coronary Care Unit)
Use of different configurations, alarm settings on the same or similar monitors in one area (e.g.,
ICU or Cardiac OR) may result in danger.
When selecting a configuration please ensure that it is appropriate for the patient being monitored.