Alarms
7-12
3) Set [Alarm Priority]:
[All]: The monitor will alarm when the other bed generates any level of alarm.
[High&Med]: The monitor will alarm when the other bed generates a high level alarm and medium level
alarm.
[High Only]: The monitor will alarm when the other bed generates a high level alarm.
7.16.3 Remote View Reset
1) Click [Main Menu] hot key → [System] → [Maintenance], enter the maintenance password and click the Enter
key.
2) Select [Alarm] tab → [Network] tab.
3) Select [Remote View Reset]. The [Alarm Reset] appear at the lower left corner of [Remote View] window. Click
[Alarm Reset] button; then Remote View alarm will be reset.
7.16.4 Alarm Reset By Remote View
User can reset the alarms occurring on the remote devices that are viewed on the Remote View screen of the
monitor. To enable this function, follow this procedure:
1) Click [Main Menu] hot key → [System] → [Maintenance], enter the maintenance password and click the Enter
key.
2) Select [Alarm] tab → [Network] → [Remote View].
3) Switch ON of OFF the [Alarm Reset By Remote View].
7.17 Set Nurse Call
If Nurse Call is active the monitor will send a request signal to the Nurse Call system for calling a nurse when the
user set alarm is triggered. The monitor provides a Nurse Call interface. After the monitor is connected with the
hospital Nurse Call system through the dedicated Nurse Call cable (available from Comen), the Nurse Call function
can be enabled.
The Nurse Call function will be triggered when the monitor meets all requirements as follows:
Nurse Call function is enabled.
The user set alarm condition is triggered.
The monitor is not in the alarm paused or reset modes.
7.17.1 Nurse Call Setup
1) Click [Main Menu] hot key → [System] → [Maintenance], enter the maintenance password and click the Enter
key.