Clinical Assistant Assessment (CAA)
17-4
17.5.2 Set the Window Time
You can follow the steps below to set the time length of the trend data displayed in each screen:
1) Go to the Graphic Trends page;
2) Set [Zoom] to:
[8min]: to display 8min trend data in each screen (you can view the trends of the last 1h);
[30min], [1h], [2h] or [4h]: to display the trend data of the selected time length in each screen (you can
view the trends of the last 4h); or
[8h], [12h], [24h] or [48h]: to display the trend data of the selected time length in each screen (you can
view the trends of the last 120h).
17.5.3 Set the Number of Waveforms
You can follow the steps below to set the number of waveforms displayed in graphic trends:
1) Go to the [Graphic Trends] page;
2) Set the value of [Wave].
17.5.4 Print the Graphic Trends Report
1) Go to the Graphic Trends page.
2) Click " " at the top left corner to go to the [Graphic Trends Report] menu;
3) Set the graphic trends report. See "
Section 21.6.4 Set the Graphic Trends Report"
.
4) In the [Graphic Trends Report] menu, click " " to print the graphic trends report.
17.6 Events Review
The Monitor can save alarm events in real time, including physiological alarm events, technical alarm events, manual
events and operational events. When an event occurs, the Monitor will save the values of corresponding parameters
at the time of the event and the 32s waveforms before and after the event.
Alarm log
Alarm log is used to record the alarm condition information, including the alarm information, priority, the associated
alarm limits and the date and time of the occurrence. All High, Medium and Low priority alarm conditions are
recorded and up to 2000 records can be stored in the system.
Event log
Event log is used to record the operations of Startup/Shutdown, change of alarm system setting, Standby, New
Patient, Calibration, System Self-test, and Alarm Audio Paused, and Alarm Paused Event.