Other functions
25-2
25.1.1 Set network protocol
1) Select [Main Menu] → [System] → [Maintenance] → input Maintenance password → click Enter
2) Select [Network Setup] tab → [Network Type] → [Network Protocol]
[CMS] :connects Comen’s Central Monitoring System.
[OEM] :connects MEDICALSYSTEM’s OEM
25.1.2 Set network type
1) Select [Main Menu] → [System] → [Maintenance] → input Maintenance password → click Enter
2) Select [Network Setup] tab → [Network Type] → [Monitor] , set your network type as: [LAN 1] or [WLAN] .
25.1.3 Set wired network
1) Click [Main Menu] hot key → [System] → [Maintenance], enter the maintenance password and click the Enter
key.
2) Select [Network Setup] tab → [LAN] tab.
Input [IP Address], [Subnet Mask] and [Gateway] manually.
25.1.4 Set wireless network
1) Click [Main Menu] hot key → [System] → [Maintenance], enter the maintenance password and click the Enter
key.
2) Select [Network Setup] tab → [WLAN] tab.
3) Select [Add WLAN] tab, add your desired WLAN, and complete the settings.
4) Select [WLAN] and set [SSID], [Security] and [Password].
5) Select [WLAN IP], input [IP Address], [Subnet Mask] and [Gateway] manually.
25.2 Central monitoring system connection
1) Set network protocol to [CMS] according to “
25.1.1 Set Network Protocol
”.
2) Set [Network Type] as [LAN 1] or [WLAN] according to “
25.1.2 Set network type
” .
3) Set the network accoding to “
25.1.3 Set wired network
” or “
25.1.4 Set wireless network
” to connect Central
Monitoring Sytem
4) Refer to Comen Central Monitoring System Instructions Manual for detailed function descriptions when the
connection to the central monitoring system is established.