C.O. Monitoring
16-3
16.5 C.O. Setup
16.5.1 Set BT Alarm
1) Select the C.O. parameter area to enter [C.O.] menu.
2) Select [Alarm] tab.
3) Set the alarm as required.
16.5.2 Set TEMP Unit
1) Select [Main Menu] hot key → [System] → [Maintenance], enter the maintenance password and click the
Enter key.
2) Select [Unit].
3) Set [TEMP Unit].
16.5.3 Set C.O. Timeout
If the C.O. measurement value is not updated again within the set time the measurement value will be displayed in
a hollow form to indicate that such value becomes invalid. Set C.O. Timeout in the following way:
1) Select [Main Menu] hot key → [System] → [Maintenance], enter the maintenance password and click the
Enter key.
2) Set [Module] tab.
3) Set [Other] tab.
4) Set [C.O. Timeout].
16.5.4 Set Patient Information
1) Select the C.O. parameter area to enter [C.O.] menu.
2) Select [Setup] tab.
3) Input [Height (cm)] and [Weight (kg)].
16.5.5 Set Interval
In order to ensure accurate measurement results, the next measurement should be carried out when the blood
temperature returns to normal, thus an interval should be set between the two measurements.
1) Select the C.O. parameter area to enter [C.O.] menu.
2) Select [Setup].