Battery
22-2
22.2 Installing the Battery
The Monitor's battery must be installed and replaced by trained and authorized maintenance personnel. The
battery is not installed when the Monitor leaves the factory. Please contact a maintenance engineer for battery
installation before using the Monitor for the first time.
22.3 View Battery Information
1) Click [Main Menu] hot key → [System] → [Maintenance], enter the maintenance password and click the Enter
key;
2) Click [Battery Information] tab.
22.4 Optimize and Check Battery Performance
22.4.1 Optimize Battery Performance
The battery must undergo at least two complete optimization cycles before first use. The battery's performance will
gradually decrease as the time of use increases. It is recommended that you optimize the battery every three
months. If it has not been optimized for at least three months the displayed battery voltage level may not be
accurate.
When optimizing the battery, please ensure the following:
1) Completely disconnect the monitor from the patient and stop all monitoring and measurement.
2) Put the battery for optimization in the battery case of the device.
3) Please ensure that the battery is charged without interruption until it is fully charged.
4) Disconnect the AC power supply and use the battery to supply voltage to the monitor until the monitor shuts
down automatically.
5) Battery optimization is now complete.
22.4.2 Check Battery Performance
The battery life varies with the storage, operational environment, frequency of battery discharging and length of
use. The battery performance will degrade gradually even if the battery is not used. A battery performance check