ECG Monitoring
8-4
8.3.3 Install ECG Leads
The table below lists the names of leads in European and American standards. (RA, LA, RL, LL and V are used to
represent leads in AHA code, whereas R, L, N, F and C are used in the European code):
American code
European code
Electrode identifier
Color code
Electrode identifier
Color code
RA
White
R
Red
LA
Black
L
Brown
RL
Green
N or RF
Black
LL
Red
F
Green
V1
Red
C1
Red
V2
Brown
C2
Brown
V3
Green
C3
Green
V4
Blue
C4
Brown
V5
Orange
C5
Black
V6
Purple
C6
Purple
8.3.3.1 Place 3-lead Electrodes
Refer to American and European standards for placing electrodes of a 3-lead unit:
White/red (right arm) electrode Place it below the clavicle, near the right arm.
Black/yellow (left arm) electrode Place it below the clavicle, near the left arm.
Red/green (left leg) electrode Place it at the left lower abdomen.