IBP Monitoring
14-7
(1) Parameter (2) Systolic pressure alarm limit (3) Systolic pressure
(4) Diastolic pressure (5) Diastolic pressure alarm limit (6) Mean Pressure
(7) Mean pressure alarm limit
14.5 IBP Setup
14.5.1 Set IBP Alarm
1) Select the IBP parameter area or waveform area to enter the corresponding IBP menu.
2) Select [Alarm] tab.
3) Set the alarm as required.
14.5.2 Set Auto Alarm Limits
1) Select the IBP parameter area or waveform area to enter the corresponding IBP menu.
2) Select [Alarm] tab.
3) Select [Auto Alm Limits], and the alarm limits will be automatically adjusted.
14.5.3 Change Pressure Label
1) Select the IBP parameter area or waveform area to enter the corresponding pressure menu.
2) Select [Setup] tab.
3) Select [IBPX Label] (X represents IBP channel), and select the appropriate label.
Label
Description
Label
Description
Art
Arterial pressure
Ao
Aortic pressure
PA
Pulmonary artery
pressure
UAP
Umbilical artery pressure
CVP
Central venous pressure
BAP
Brachial artery pressure
RAP
Right atrial pressure
FAP
Femoral artery pressure
LAP
Left atrial pressure
UVP
Umbilical venous
pressure
ICP
Intracranial pressure
IAP
Intra-abdominal pressure
P1/P2
Expansion pressure
P3/ P4
Expansion pressure
(5)
(3)
(2)
(4)
(6)
(7)
(1)