EasyManua.ls Logo

Comen N10 - Page 249

Default Icon
307 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Product Specifications
II-3
Memory cardstandard capacity
Trend data
120h
NIBP measurement
1000 groups
Alarm Event
1000 groups
Waveform length
48h
Arrhythmia Statistics
48h
OxyCRG events
400 events
12-lead ECG analysis
20 groups
ST review
120h
Memory cardlarge capacity
Trend data
240h
NIBP measurement
3500 groups
Alarm Event
2500 groups
Waveform length
48h
Arrhythmia Statistics
48h
OxyCRG events
400 events
12-lead ECG analysis
20 groups
ST review
120h
6) Record specifications
Name
Specification
Record Paper Width
50mm
Print width
48mm
Paper speed
12.5mm/s, 25 mm/s, 50mm/s
Record time
8s, 16s, 32s, Continual
No. of Waveform
Up to 3 curves
Timed Record Interval
10min, 20min, 30min, 40min, 50min, 1h, 2h, 3h, 4h and Off
7) ECG specifications
Name
Specification
Standard;
IEC 60601-2-27:2011/IEC 60601-2-25:2011
Lead Type
3-Lead; 5-Lead; 6-Lead, 12-Lead and Auto
Overload Protection
Loaded with differential input-circuit voltages of 1 V peak-to-valley, no damage
Lead off detection and noise
suppression current
Measurement electrode: <0.1µA
Driven electrode: <900nA
Auto identification of ECG lead type
QRS wave amplitude and
Amplitude (p-v RTI) range
0.5mV~5mV

Table of Contents