Product Specifications
II-14
Systole pressure (SYS), Diastolic pressure (DIA), Mean pressure (MAP)
Pulse Rate (PR) in NIBP list
Measurement
range in ADU mode
25mmHg - 290mmHg (3.33kPa-38.7kPa)
10mmHg - 250mmHg (1.3kPa-33.3kPa)
15mmHg - 260mmHg (2.0kPa-34.7kPa)
Measurement
range in PED. mode
25mmHg - 250mmHg (3.33kPa-33.3kPa)
10mmHg - 200mmHg (1.3kPa-28.0kPa)
15mmHg - 225mmHg (2.0kPa-30.0kPa)
Measurement
range in NEO.
mode
25mmHg - 140mmHg (3.33kPa-18.7kPa)
10mmHg - 115mmHg (1.3kPa-15.3kPa)
15mmHg - 125mmHg (2.0kPa-16.7kPa)
Static pressure
measurement range and
accuracy
Measurement range: 0mmHg - 300mmHg (0 kPa - 40.0 kPa);
Measurement accuracy: ±3mmHg (±0.4 kPa)
Max. mean difference: ±5mmHg
Max. standard deviation: 8mmHg
Initial inflation pressure
range
a) In Adu. mode: 80mmHg - 290mmHg (10.7kPa - 38.7kPa); default: 160mmHg
b) In Ped. mode: 80mmHg - 240mmHg (10.7kPa - 32.0kPa); default: 120mmHg
c) In Neo. mode: 60mmHg - 140mmHg (8.0kPa - 18.7kPa); default: 100mmHg
Software Over pressure
protection range and
accuracy
297mmHg ±3mmHg (39.6kPa±0.4kPa)
297mmHg ±3mmHg (39.6kPa±0.4kPa)
147mmHg ±3mmHg (19.6kPa±0.4kPa)
120s in Adu./Ped. mode;
85s in Neo. mode
125s in Adu/Ped. Mode;
87s in Neo. mode
Alarm setting rules:
a) 4 types of alarm limits, including extreme high limit, upper limit, lower limit, and
extreme low limit.
b) Step:
10mmHg - 50mmHg: 1mmHg.
51mmHg - 290mmHg: 5mmHg
Extreme high limit set range: 28mmHg-290mmHg