Product Specifications
II-16
Extreme low limit set range: 15mmHg - 110mmHg
Blood measurement mode
Manual, auto (periodic), sequence, hourly and continuous mode
Measuring interval in
auto mode
1min, 2min, 2.5min, 3min, 4 min, 5 min, 10 min,
15 min, 30 min, 60 min, 90min, 120 min, 180 min, 240
min, 480 min, 720min
Continuous mode cycling
time
5 min.
SunTech NIBP
NIBP measurement range
Measurement
range in Adu. mode
SYS
40mmHg - 260mmHg (5.3kPa - 34.7kPa)
DIA
20mmHg - 200mmHg (2.7kPa - 26.7kPa)
MAP
26mmHg - 220mmHg (3.5kPa - 29.3kPa)
Measurement
range in Ped. mode
SYS
40mmHg - 160mmHg (5.3kPa - 21.3kPa)
DIA
20mmHg - 120mmHg (2.7kPa - 16.0kPa)
MAP
26mmHg - 133mmHg (3.5kPa - 17.7kPa)
Measurement
range in Neo. mode
SYS
40mmHg - 130mmHg (5.3kPa - 17.3kPa)
DIA
20mmHg - 100mmHg (2.7kPa - 13.3kPa)
MAP
26mmHg - 110mmHg (3.5kPa - 14.7kPa)
Static pressure
measurement range and
accuracy
Measurement range: 0mmHg - 300mmHg (0 kPa - 40.0 kPa);
Measurement accuracy: ±3mmHg 0.4 kPa)
Measurement Error
Max. mean difference: ±5mmHg
Max. standard deviation: 8mmHg
Initial inflation pressure
range
a) In Adu. mode: 120mmHg - 280mmHg (16.0kPa - 37.3kPa); default: 160mmHg
b) In Ped. mode: 80mmHg - 170mmHg (10.7kPa - 22.7kPa); default: 120mmHg
c) In Neo. mode: 60mmHg - 140mmHg (8.0kPa - 18.7kPa); default: 100mmHg
Software Over pressure
protection range and
accuracy
Adu. mode
297mmHg ±3mmHg (39.6kPa ±0.4kPa)
Ped. mode
297mmHg ±3mmHg (39.6kPa ±0.4kPa)
Neo. mode
147mmHg ±3mmHg (19.6kPa ±0.4kPa)
Max. measuring period
130s in Adu. mode;
90s in Ped. mode;
75s in Neo. mode
Max. venipuncture time
125s in Adu/Ped. Mode;
87s in Neo. mode
NIBP alarm limit setting
Alarm setting rules: