EMC
V-6
Guidance and manufacturer’s declaration electromagnetic emission
N10/N10A/N10C/N12/N12A/N12C/N15/N15A/N15C monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of NC8/NC8A/NC10/NC10A/NC12/NC12Amonitor
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
Conducted and radiated RF
emissions
CISPR 11
Group 1
N10/N10A/N10C/N12/N12A/N12C/N15/N15A/N15C
monitor uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
Conducted and radiated RF
emissions
CISPR 11
Class A
N10/N10A/N10C/N12/N12A/N12C/N15/N15A/N15C
monitor is suitable for use in all establishments other
than domestic ones and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic distortion
emissions IEC 61000-3-2-
2014
Pass
Voltage fluctuations/
flicker emissions
IEC 61000-3-3-2013
Pass
Guide and Manufacturer’s Statement – Electromagnetic Immunity
N10/N10A/N10C/N12/N12A/N12C/N15/N15A/N15C monitor is intended for use in the electromagnetic
environment specified below. The customer or the user of NC8/NC8A/NC10/NC10A/NC12/NC12Amonitor
should assure that it is used in such an environment.
Immunity Test
IEC60601 Test Level
Compliance Level
Electromagnetic
Environment Guide
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV
air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15
kV air
The relative humidity
should be at least 30 %.
Electrical fast transient
burst (EFT/B)
IEC 61000-4-4
± 2 kV for input a.c. and
d.c. power input port
± 1 kV for SIP/SOP port
100 kHz repetition
frequency
± 2 kV for input a.c. and
d.c. power input port
100 kHz repetition
frequency
Connect the equipment
to supply mains of
professional healthcare
facility other than
public supply mains
SIP/SOPS whose
maximum cable length
is less than 3 m in length
are excluded
Surge
IEC 61000-4-5
± 0.5kV, ± 1 kV line to line
± 0.5kV, ± 1 kV, ± 2 kV line
to ground
± 0.5kV, ± 1 kV line to
line
± 0.5kV, ± 1 kV, ± 2 kV line
to ground
Connect the equipment
to supply mains of
professional healthcare
facility other than public
supply mains