6 Operating mode principles
Shut down system
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
6 - 5
Regular shut down
► Switch the circulation off.
► Switch analysers off.
► Close all media connections.
► Depressurise the system.
► Turn the main switch from the “I” position (on) to the “0” position (off).