11 Troubleshooting
Eliminating faults
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
11 - 3
11.1.4 Antechamber
Fault Debug
Power failure
(Siemens PLC-controlled
antechambers)
If power is available:
Refill antechamber.
If no power is available:
Eliminate vacuum by loosing a flange on the pip-
ing outside the antechamber.
Unexpected pressure in-
crease or drop
Check connections of working gas / purge gas
and vacuum pump.
Check seals.
Perform a leak test.
Antechamber door mechan-
ism jammed
Risk of damage to the locking mechanism:
Do not use force to open.
If there is a vacuum and it must be eliminated to open
the door:
Loosen screws on the flange below the antecham-
ber.
Antechamber door mechan-
ism jammed
If the door is locked:
Check whether the “door locked” status is dis-
played on the touch panel.
Check for vacuum in the antechamber.
If the fault cannot be eliminated with the debug possibilities in the table above and there
is vacuum or overpressure in the antechamber:
WARNING
Inert gas atmosphere escapes into the room
Personal injury
Conduct a risk analysis with regard to toxic gases inside the glovebox or enclosure
and the requirement of a safe transfer out (and in) procedure with external purge or
refill gas before starting the transfer out sequence.
In doubt use personal protection equipment (PPE).
Eliminate vacuum by loosing a flange on the piping outside the antechamber.