9 Antechamber operation (Vacuum antechamber)
Optional antechamber features
Page
9 - 20
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Antechamber cycle Display on touch panel
¡ Inner door is opened
the first time
¡ Outer door is locked
immediately
¡ Inner door is closed
¡ Outer door remains
locked
9.7 Optional antechamber features
MBRAUN product Description MBRAUN sales number
MB-SCH-PLC-2-DV Siemens PLC-controlled ante-
chamber door interlock
Prevents a simultaneous re-
lease of the inner and outer
doors.
1500284
MB-AC-purge (PLC) Purge antechamber from ex-
ternal source (loading), Siemens
PLC-controlled:
¡ Transfer of vacuum-sensit-
ive materials.
¡ Purge of the antechamber
with inert gas from external
source or with ambient air if
hazardous material is used.
¡ Inertisation of the antecham-
ber if vacuum-sensitive
materials are loaded / un-
loaded.
1502054
MB-AC-Purge (manual) Purge antechamber from ex-
ternal source (loading):
¡ Transfer of vacuum-sensit-
ive materials.
¡ Purge of the antechamber
with inert gas from external
source or with ambient air if
hazardous material is used.
¡ Inertisation of the antecham-
ber if vacuum-sensitive
materials are loaded / un-
loaded.