4 Transport, installation and commissioning
Installation
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
4 - 11
4.2.1 Definition of the clean room volume
WARNING
Inert gas of high (or infinite) amount
Inert gas may be released into the environment as a result of malfunctions or faulty oper-
ation. This creates a suffocation risk in closed rooms, especially if the fault occurs in the
absence of any personnel.
► Make sure that the automatic switch-off function parameter is activated.
► If automatic switch-off is deactivated, be sure that the room is ventilated to prevent
the O
2
concentration in the ambient air from falling below 19%.
► Always ensure sufficient room ventilation.
► Connect the purge exhaust of the inert gas system to a sufficient powerful on-site ex-
haust system. For the required exhaust system flow rate refer to the technical data of
the MBRAUN system media connections.
Example: Combination with a MBRAUN glovebox
1 Overall volume of the room 2 Furnitures and other equipment inside
the room
3 Workstation or
Glovebox with gas purifier
3a Glovebox