12 Maintenance
Introduction
Page
12 - 2
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
12.1.2 Spare parts kit
In order to prevent production breakdowns, you should keep spare parts on hand for un-
planned maintenance work so that they are readily available (there is a spare part kit per
system available or single spare parts like gloves, analysers, can be ordered – refer to
“Spare parts”).
12.1.3 Who may do what?
Skilled personal
(mechanical)
Skilled electricians or
technicians
MBRAUN Service
only
Exchange/repair of
electrical compon-
ents
X
Exchange of elec-
trical-pneumatic
components
X (24V) X
Repair of pressur-
ized pipework
X
Maintenance work
according to oper-
ating manual
X
Change of the mol
sieve filter or cop-
per catalyst
X
Calibration work on
the system or
sensors
X
Change of the ac-
tivated carbon fil-
ters
X
12.1.4 Qualification of maintenance personnel
WARNING
Insufficient qualified personnel
Risk of injuries
► Ensure that maintenance work is done only by personnel, qualified for the specific
task.
► Ensure that the maintenance personnel is trained in operating an inert gas system
and has sufficient experience with glovebox / enclosure purge function
► Ensure that the maintenance personnel is experienced in operation of systems with
high inert gas volumes