9 Antechamber operation (Vacuum antechamber)
PLC-controlled loading and unloading
Page
9 - 12
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Touch button to start or stop vacuum pump:
Single vacuum pump is used for glovebox operation and
for antechamber operation
9.5.2 Parameters
Parameters for time controlled AC operation
Time controlled antechamber: antechamber parameters are set on the Antechamber
Parameter screen
Start Screen Antechamber icon Parameter
Parameters for Siemens
PLC-controlled antecham-
ber operation
16
Parameter Explanation
Number of cycles Number of refilling cycles
Evacuation time Evacuation time for each cycle
Time between refills Refill time for each cycle
Time to final refilling Refill time for last refill cycle
Parameters for pressure controlled AC operation
Vacuum controlled antechamber: antechamber parameters are set on the Antechamber
Parameter screen
16
For preset values refer to parameter table below in chapter Human machine interface