12 Maintenance
Maintenance plan overview
Page
12 - 4
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
12.2.3 Regenerative LMF
T: Test
C: Clean
R: Repair or replace
UNIT
Monthly
Quarterly
Semi-annually
Annually
Others
Replace the mol sieve filling when its life
time is about to expire
(R)
3-5 years
Magnetic valve VRA (R)
as
needed
12.2.4 Gloves
T: Test
C: Clean
R: Repair or replace
UNIT
Monthly
Quarterly
Semi-annually
Annually
Others
Check gloves for damage T
Replace gloves and fixation O-rings (R)
2 years
at the
latest
Gloveport outer surface C
Gloveport tightness and inner / outer O-
ring to the window in combination with
leak test (typically Argon test)
T (R) *)
as
needed
12.2.5 Mini Antechamber
T: Test
C: Clean
R: Repair or replace
UNIT
Monthly
Quarterly
Semi-annually
Annually
Others
Check door sealing surface T