8 Glovebox or enclosure operation
Additional function and equipment from MBRAUN
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
8 - 17
Example:
Trigger value = 100 ppm
Automatic purge on at an O
2
concentration of ≥ 100 ppm
Automatic purge off at an O
2
concentration of ≤ 90 ppm
Prerequisite for an automatic glovebox purge
¡ Parameter in Parameter Box Purging screen must be configured correctly.
¡ Circulation must be switched on.
¡ Analysers must be switched on and work properly.
Automatic activation of the glovebox purge: if the specified trigger value is exceeded
Automatic deactivation the glovebox purge: if the actual value is less than the specified
trigger value
Function
Status Process initiated by automatic box purge
function
Trigger value is exceeded ¡ Circulation is stopped
¡ Purge of the box is started
¡ Reference value is monitored
Actual value less than trigger value ¡ Purge gas input and exhaust valves
are closed
¡ Circulation starts automatically
8.11 Additional function and equipment from MBRAUN
8.11.1 Additional equipment
For more information ask MBRAUN sales.
Glovebox atmo-
sphere
Particle absorber Suction removal of particles emitted
from processes in the box
Differential pressure
sensor
Monitoring the of box particle filter sat-
uration
UGP Universal measurement device for vari-
ous gases/solvent vapours
GSU Determination of a specific gas mixture
(H
2
O/ O
2
)