7 Gas purifier operation
Regeneration operation
Page
7 - 18
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
7.5.2 Reactor regeneration
Manual activation / deactivating of a reactor regeneration
Prerequisites
¡ A sufficient quantity of regeneration gas is available.
¡ The reactor that is to be regenerated is out of circulation.
¡ The vacuum pump is activated.
¡ No manual antechamber cycle is allowed during the subsequent regeneration period.
¡ Condensate is trapped into a bottle or the like.
Activating regeneration (example double purifier, dp)
Start Screen ⇨ Functions
Reactor 2 is to be regen-
erated:
► Touch the Regener-
ation Reactor 2 but-
ton.
ð A confirmation dia-
logue 1 is displayed
(see figure below).
Reactor 1 is to be regen-
erated:
► Touch the Regener-
ation Reactor 1 but-
ton.
ð The system reaction
is like reactor 1 regen-
eration.
Confirmation dialogue 1:
► Touch “Yes”.
ð Confirmation dialogue1 disappears.
ð Confirmation dialogue2 is displayed.
Confirmation dialogue2