12 Maintenance
Maintenance plan overview
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
12 - 5
Door sealing R
O-ring angle valves R
12.2.6 Main Antechamber
T: Test
C: Clean
R: Repair or replace
UNIT
Monthly
Quarterly
Semi-annually
Annually
Others
Check for smooth operation of the guide
rails for the tray
T (R)
as
needed
Check for leaks from the gas strut
Check smooth operation and for uninter-
rupted movement
T (R)
as
needed
Check to make sure that all connections
are tight and free of damage
T
Check antechamber door seals for dam-
age
T
Replace sealing rings
(replace if damaged)
R
Optional:
Safety valve test 0.1bar / 0.3bar
1
2
T
Optional:
MBRAUN overpressure safety valve
(OSV)
10-15mbar (test is described in the com-
ponent manual)
T (R)
as
needed
1
See also maintenance advices from 3rd party.