7A ECO Mode
Operation
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
7A - 5
Parameter "Automated Activation ECO-Mode":
YES ECO Mode starts at the time set in the parameter "Time of Day for Autom.
Activation"
NO ECO Mode is switched off
Reduce blower speed
Parameter "Blower Speed Reduction":
YES Blower speed is reduced to the value set in parameter "Reduced Speed"
NO Blower speed is not reduced
► Enter the value for the reduced speed (input range 60 % … 100 %)
7A.4.3 Reduced Blower speed
Automatic control of the atmosphere quality
If the O
2
or H
2
O concentration exceeds the value set in parameter "Max. O2-conc. At
Speed Reduction (ppm)", the blower speeds up automatically to 100% to increase the
quality of the box atmosphere.
The threshold value for the O
2
and H
2
O concentration must be lower than the threshold
for activating the automatic box purge.