2 Inert gas technology
Components
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
2 - 13
2.2.5 Vacuum pump
Standard vacuum pumps
In MBRAUN standard systems, all components use the vacuum pump of the gas purifier
as a shared vacuum pump.
The vacuum pump supports several system functions:
¡ Glovebox/enclosure pressure regulation
¡ Antechamber operation
¡ Regeneration of a reactor
¡ Regeneration of an optional solvent filter
If one of these functions is frequently in use, MBRAUN recommends adding an additional
vacuum pump, dedicated to the frequently used function.
For information about requirements, please contact MBRAUN sales.
The vacuum pump is a 3
rd
- party component. Refer to the 3
rd
- party documentation for
more information.
Dry vacuum pump
1 Gas inlet NW25
2 Gas outlet (exhaust) NW25
3 Gas ballast control
Oil sealed vacuum pump
1 Oil mist filter (on outlet port)
2 Inlet hose
3 Gas-ballast inlet control (adjusted in the fact-
ory)
Additional vacuum pumps
Additional vacuum pumps can be used for special purposes, such as:
¡ Special antechamber processes (antechamber oven)
¡ Solvent filters
¡ Other components