7A ECO Mode
Operation
Page
7A - 8
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
7A.4.7 Box light
Parameter "Switch off Box-light":
YES Glovebox light is switched off on ECO-
Mode activation
NO Box light remains on during ECO Mode
operation
The box light can be switched on and off manually in
ECO Mode.
If a multi-glovebox operation environment is used
and only one glovebox is to remain lit during ECO
Mode operation:
► Set the parameter to NO
ð Light stays on
► Switch off the light at the desired box using the
hardware switch, located directly on the fluores-
cent lamp (refer to figure)
1 Light is on 2 Light is off