12 Maintenance
Maintenance plan overview
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
12 - 3
MBRAUN offers training for your personnel to perform maintenance work on purification
systems.
12.2 Maintenance plan overview
The specifications of the intervals are minimum specifications. Depending on the pro-
cesses and process materials used, shorter intervals may be necessary.
*) Work done only by MBRAUN Service personal
**) work done only by trained HVAC technicians
***) work done only by trained electricians
12.2.1 System in general
T: Test
C: Clean
R: Repair or replace
UNIT
Monthly
Quarterly
Semi-annually
Annually
Others
Leak test T T
Electrical test
(German DGUVV3, annual E-Check for
safety)
T
Pressure tests/ leak tests on pressurized
pipework which have to be tested regu-
larly acc. to national rules (e.g. pressure
device directive)
T (T)
2years
12.2.2 Non-regenerative LMF
T: Test
C: Clean
R: Repair or replace
UNIT
Monthly
Quarterly
Semi-annually
Annually
Others
Replace filter medium (activated carbon;
interval depends on process)
R (R)
depend-
ing on
solvent
load
(SOP)